Product Introduction
The ZHK2503N gas flowmeter consists of a flowmeter module with an orifice, a soap film flowmeter module, a mass flowmeter module, a microcontroller, a display screen, and other components.
Get a quoteMulti-use: The main unit can be used with five different ranges of soap film flowmeters, seven different ranges of orifice flowmeters, and three different ranges of mass flowmeters.
Touchscreen Color Display: User-friendly interface and simple operation.
Automatic Display of Conditions: Can automatically display volumetric flow rates under four conditions: current conditions, standard conditions (273.15K, 101.325kPa), standard conditions (293.15K, 101.325kPa), and standard conditions (298.15K, 101.325kPa).
Digital Calibration Capability: Capable of digital calibration for flow, pressure, environmental temperature, time values, and battery level.
Data Storage Function: Can store 80 data points per range.
USB Interface: Option to add a USB drive for data export.
Bluetooth Printer Option: Supports wireless communication between the printer and the main unit.
Manual Input Capability: Allows manual input of standard atmospheric pressure and environmental temperature values to minimize measurement errors caused by automatic sensor measurement. Built-in Lithium Battery: Battery life exceeds 8 hours. Battery Level Display and Over-discharge Protection.
Soap Film Section: The soap film flow detector utilizes an infrared sensing device combined with microcomputer data processing technology, ensuring high measurement accuracy and effectively reducing human errors.
The soap film tube has good light shielding properties, preventing sunlight interference and ensuring high reliability.
Capable of automated calibration of soap film volume values.
Can automatically record and display the flow values for N measurements (unlimited times), and automatically calculate the average flow of the N measurements.
Orifice Section: Utilizes a high-precision differential pressure sensor for real-time measurement and display of gas flow.
Application
Suitable for flow calibration and verification of large, medium, and small flow particulate samplers, atmospheric/smoke/dust samplers, VOC samplers, gas chromatography, float flowmeters, mass flowmeters, and other instruments.
Mass Flowmeter Section: Utilizes a high-precision mass flow sensor for real-time measurement and display of gas flow.