Application
Suitable for collecting TSP / PM10 / PM2.5 / radioactive aerosol samples (cutting devices can be optional).
Get a quoteMain Features
Utilizes high-precision pressure, temperature, and flow sensors along with microcontroller measurement and control technology to automatically control the flow rate of the suction pump, achieving constant flow sampling.
Automatically measures environmental temperature, atmospheric pressure, pre-sampling temperature, pre-sampling pressure, flow rate, and other parameters.
Digitally sets sampling flow rate (working conditions), sampling time, interval sampling time, sampling volume, and other parameters.
Digitally calibrates values for flow, temperature, pressure, time, and other metrics.
Automatically displays values for flow rate, sampling volume, temperature, pressure, sampling time, power outage duration, and other metrics.
Overload automatic protection: If the sampling flow deviates from the set flow rate by more than ±10% for 5 consecutive minutes during sampling, the instrument automatically shuts down for protection.
Power outage protection function: If a power outage occurs during sampling, the device can automatically save the sampling data and resume sampling once power is restored.
Capable of storing 100 sets of sampling data and power outage data.
Uses a brushless fan, which has a long service life and can operate without failure for 10,000 hours.
The system is well-integrated, featuring a convenient assembly and disas-sembly method, along with effective protective measures.
PM10 and PM2.5 cutting devices come with a test certificate.